???displayArticle.abstract???
We present a new technique for fabricating planar patch electrodes in the laboratory. Planar electrodes are micromolded using a micron-sized stream of air to define an aperture in the silicone elastomer, polydimethylsiloxane (PDMS). We have previously demonstrated that planar PDMS electrodes make excellent patch electrodes after surface modification. We demonstrate single-channel measurements of the rSlo channel in Xenopus oocytes and whole-cell measurements in CHO and RBL mammalian cell lines, using planar PDMS electrodes.
Armstrong,
Inactivation of the sodium channel. II. Gating current experiments.
1977, Pubmed
Armstrong,
Inactivation of the sodium channel. II. Gating current experiments.
1977,
Pubmed Brueggemann,
Ion channel drug discovery and research: the automated Nano-Patch-Clamp technology.
2004,
Pubmed Díaz,
Role of the S4 segment in a voltage-dependent calcium-sensitive potassium (hSlo) channel.
1998,
Pubmed
,
Xenbase Hamill,
Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches.
1981,
Pubmed Kiss,
High throughput ion-channel pharmacology: planar-array-based voltage clamp.
2003,
Pubmed Klemic,
Micromolded PDMS planar electrode allows patch clamp electrical recordings from cells.
2002,
Pubmed
,
Xenbase Lee,
Solvent compatibility of poly(dimethylsiloxane)-based microfluidic devices.
2003,
Pubmed Lindau,
A patch-clamp study of histamine-secreting cells.
1986,
Pubmed Methfessel,
Patch clamp measurements on Xenopus laevis oocytes: currents through endogenous channels and implanted acetylcholine receptor and sodium channels.
1986,
Pubmed
,
Xenbase Pantoja,
Bilayer reconstitution of voltage-dependent ion channels using a microfabricated silicon chip.
2001,
Pubmed Schmidt,
A Chip-Based Biosensor for the Functional Analysis of Single Ion Channels We thank E. Ermanntraut, L. Giovangrandi, T. Wohland, A. Brecht, M. Köhler, C. Bieri, D. Stamou, and R. Hovius for advice. This work was supported by the Swiss National Science Foundation (Priority Program for Biotechnology) and by an interdepartmental grant of the Swiss Federal Institute of Technology Lausanne (EPFL, Project Microtechnique 96).
2000,
Pubmed Sigworth,
Patch clamp on a chip.
2002,
Pubmed Stühmer,
Electrophysiologic recordings from Xenopus oocytes.
1998,
Pubmed
,
Xenbase Xu,
Ion-channel assay technologies: quo vadis?
2001,
Pubmed Xu,
A benchmark study with sealchip planar patch-clamp technology.
2003,
Pubmed